  • Switch off TACE-II+ equipment. Press the "Value" & "Menu" switch at once & Switch on the TACE-II+ & hold the pressing until the LCD screen of TACII++ display the  Scale Calibration.
  • After displaying the "Scale Calibration", Screen will instruct "Remove all bags & press "START". Then press "START" button.
  • After that It will instruct to put 200gm weight on plasma scale. Then do accordingly. TACE-II+ will measure 200gm on plasma scale & will calibrate that.
  • After that It will instruct to put 200gm weight on SAGM scale. Then do accordingly. TACE||+ will measure 200gm on SAGM & will calibrate that.
  • After that It will instruct to put 200gm weight on BC scale. Then do accordingly. TACE-II+ will measure 200gm on BC & will calibrate that.
  • After calibrating all scales, TACEII+ will ask for save setting for recent calibration date. Then change the date by "VALUE" button to current date/month from the previous calibrated date. Use the "shift" key for change the parameter of the setting like date/month/year.
  • After saving the final setting for calibration press "STOP" for save the final Calibration setting.

     

SOP forSoftware Calibration of T-ACE-II+
  • Switch On UPS & Air Compressor.
  • After 10 minutes Turn on T-ACE.
  • Select program from 1 to 10.
  • 4) Complete separation as required.
  • After separation close the door without any bag, wait until T-ACE shows Head Scale – "Push Start" – then switch off the machine.
  • Do not switch off machine as any other position. Follow above step only.